NIOSH Manual of Analytical Methods (NMAM), Fourth Edition
PARTICULATES NOT OTHERWISE REGULATED, RESPIRABLE 0600
DEFINITION: aerosol collected by sampler
with 4-µm median cut point
CAS: None RTECS: None
METHOD: 0600, Issue 3 EVALUATION: FULL Issue 1: 15 February 1984
Issue 3: 15 January 1998
OSHA: 5 mg/m³
NIOSH: no REL
ACGIH: 3 mg/m³
PROPERTIES: contains no asbestos and quartz less than 1%;
penetrates non-ciliated portions of respira-
tory system
SYNONYMS: nuisance dusts; particulates not otherwise classied
SAMPLING
SAMPLER: CYCLONE + FILTER (10-mm nylon cyclone,
Higgins-Dewell [HD] cyclone, or aluminum
cyclone + tared 5-µm PVC membrane)
FLOW RATE: nylon cyclone: 1.7 L/min
HD cyclone: 2.2 L/min
Al cyclone: 2.5 L/min
VOL-MIN: 20 L @ 5 mg/m³
-MAX: 400 L
SHIPMENT: routine
SAMPLE
STABILITY: stable
BLANKS: 2 to 10 eld blanks per set
ACCURACY
RANGE STUDIED: 0.5 to 10 mg/m³ (lab and eld)
BIAS: dependent on dust size distribution [1]
OVERALL
PRECISION (
): dependent on size distribution [1,2]
ACCURACY: dependent on size distribution [1]
MEASUREMENT
TECHNIQUE: GRAVIMETRIC (FILTER WEIGHT)
ANALYTE: mass of respirable dust fraction
BALANCE: 0.001 mg sensitivity; use same balance
before and after sample collection
CALIBRATION: National Institute of Standards and
Technology Class S-1.1 or ASTM Class 1
weights
RANGE: 0.1 to 2 mg per sample
ESTIMATED LOD: 0.03 mg per sample
PRECISION: <10 µg with 0.001 mg sensitivity balance;
<70 µg with 0.01 mg sensitivity balance
[3]
APPLICABILITY: The working range is 0.5 to 10 mg/m³ for a 200-L air sample. The method measures the mass concentration
of any non-volatile respirable dust. In addition to inert dusts [4], the method has been recommended for respirable coal
dust. The method is biased in light of the recently adopted international denition of respirable dust, e.g., ≈ +7% bias for
non-diesel, coal mine dust [5].
INTERFERENCES: Larger than respirable particles (over 10 µm) have been found in some cases by microscopic analysis
of cyclone lters. Over-sized particles in samples are known to be caused by inverting the cyclone assembly. Heavy dust
loadings, bers, and water-saturated dusts also interfere with the cyclones size-selective properties. The use of conductive
samplers is recommended to minimize particle charge eects.
OTHER METHODS: This method is based on and replaces Sampling Data Sheet #29.02 [6].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition
PARTICULATES NOT OTHERWISE REGULATED, RESPIRABLE: METHOD 0600, Issue 3, dated 15 January 1998 - Page 2 of 6
EQUIPMENT:
1. Sampler:
a. Filter: 5.0-µm pore size, polyvinyl chloride lter or equivalent hydrophobic membrane lter
supported by a cassette lter holder (preferably conductive).
b. Cyclone: 10-mm nylon (Mine Safety Appliance Co., Instrument Division, P. O. Box 427, Pittsburgh,
PA 15230), Higgins-Dewell (BGI Inc., 58 Guinan St., Waltham, MA 02154) [7], aluminum cyclone
(SKC Inc., 863 Valley View Road, Eighty Four, PA 15330), or equivalent.
2. Personal sampling pump, 1.7 L/min ± 5% for nylon cyclone, 2.2 L/min ± 5% for HD cyclone, or 2.5
L/min ± 5% for the Al cyclone with exible connecting tubing.
NOTE: Pulsation in the pump ow must be within ± 20% of the mean ow.
3. Balance, analytical, with sensitivity of 0.001 mg.
4. Weights, NIST Class S-1.1, or ASTM Class 1.
5. Static neutralizer, e.g., Po-210; replace nine months after the production date.
6. Forceps (preferably nylon).
7. Environmental chamber or room for balance, e.g., 20 °C ± 1 °C and 50% ± 5% RH.
SPECIAL PRECAUTIONS: None.
PREPARATION OF SAMPLERS BEFORE SAMPLING:
1. Equilibrate the lters in an environmentally controlled weighing area or chamber for at least 2 h.
2. Weigh the lters in an environmentally controlled area or chamber. Record the lter tare weight,
(mg).
a. Zero the balance before each weighing.
b. Handle the lter with forceps (nylon forceps if further analyses will be done).
c. Pass the lter over an anti-static radiation source. Repeat this step if lter does not release easily
from the forceps or if lter attracts balance pan. Static electricity can cause erroneous weight
readings.
3. Assemble the lters in the lter cassettes and close rmly so that leakage around the lter will not
occur. Place a plug in each opening of the lter cassette.
4. Remove the cyclones grit cap before use and inspect the cyclone interior. If the inside is visibly
scored, discard this cyclone since the dust separation characteristics of the cyclone may be altered.
Clean the interior of the cyclone to prevent reentrainment of large particles.
5. Assemble the sampler head. Check alignment of lter holder and cyclone in the sampling head to
prevent leakage.
SAMPLING:
6. Calibrate each personal sampling pump to the appropriate ow rate with a representative sampler
in line.
NOTE 1: Because of their inlet designs, nylon and aluminum cyclones are calibrated within a large
vessel with inlet and outlet ports. The inlet is connected to a calibrator (e.g., a bubble meter).
The cyclone outlet is connected to the outlet port within the vessel, and the vessel outlet is
attached to the pump. See APPENDIX for alternate calibration procedure. (The calibrator can
be connected directly to the HD cyclone.)
NOTE 2: Even if the ow rate shifts by a known amount between calibration and use, the nominal
ow rates are used for concentration calculation because of a self-correction feature of the
cyclones.
7. Sample 45 min to 8 h. Do not exceed 2 mg dust loading on the lter. Take 2 to 4 replicate samples for
each batch of eld samples for quality assurance on the sampling procedure (see Step 10).
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition
PARTICULATES NOT OTHERWISE REGULATED, RESPIRABLE: METHOD 0600, Issue 3, dated 15 January 1998 - Page 3 of 6
NOTE : Do not allow the sampler assembly to be inverted at any time. Turning the cyclone to
anything more than a horizontal orientation may deposit oversized material from the cyclone
body onto the lter.
SAMPLE PREPARATION:
8. Remove the top and bottom plugs from the lter cassette. Equilibrate for at least 2 h in an
environmentally controlled area or chamber.
CALIBRATION AND QUALITY CONTROL:
9. Zero the microbalance before all weighings. Use the same microbalance for weighing lters
before and after sample collection. Calibrate the balance with National Institute of Standards and
Technology Class S-1.1 or ASTM Class 1 weights.
10. The set of replicate eld samples should be exposed to the same dust environment, either in a
laboratory dust chamber [8] or in the eld [9]. The quality control samples must be taken with the
same equipment, procedures, and personnel used in the routine eld samples. Calculate precision
from these replicates and record relative standard deviation (
) on control charts. Take corrective
action when the precision is out of control [8].
MEASUREMENT:
11. Weigh each lter, including eld blanks. Record this post-sampling weight, (mg), beside its
corresponding tare weight. Record anything remarkable about a lter (e.g., visible particles,
overloading, leakage, wet, torn, etc.).
CALCULATIONS:
12. Calculate the concentration of respirable particulate, (mg/m³), in the air volume sampled, (L):
, mg/m³,
where: = tare weight of lter before sampling (mg),
= post-sampling weight of sample-containing lter (mg),
= mean tare weight of blank lters (mg),
= mean post-sampling weight of blank lters (mg),
= volume as sampled at the nominal ow rate (i.e., 1.7 L/min or 2.2 L/min).
EVALUATION OF METHOD:
1. Bias: In respirable dust measurements, the bias in a sample is calculated relative to the appropriate
respirable dust convention. The theory for calculating bias was developed by Bartley and Breuer
[10]. For this method, the bias, therefore, depends on the international convention for respirable
dust, the cyclones’ penetration curves, and the size distribution of the ambient dust. Based on
measured penetration curves for non-pulsating ow [1], the bias in this method is shown in Figure 1.
For dust size distributions in the shaded region, the bias in this method lies within the ± 0.10
criterion established by NIOSH for method validation. Bias larger than ± 0.10 would, therefore, be
expected for some workplace aerosols. However, bias within ± 0.20 would be expected for dusts
with geometric standard deviations greater than 2.0, which is the case in most workplaces.
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition
PARTICULATES NOT OTHERWISE REGULATED, RESPIRABLE: METHOD 0600, Issue 3, dated 15 January 1998 - Page 4 of 6
Bias can also be caused in a cyclone by the pulsation of the personal sampling pump. Bartley, et
al. [12] showed that cyclone samples with pulsating ow can have negative bias as large as −0.22
relative to samples with steady ow. The magnitude of the bias depends on the amplitude of the
pulsation at the cyclone aperture and the dust size distribution. For pumps with instantaneous
ow rates within 20% of the mean, the pulsation bias magnitude is less than 0.02 for most dust size
distributions encountered in the workplace.
Electric charges on the dust and the cyclone will also cause bias. Briant and Moss [13] have found
electrostatic biases as large as −50%, and show that cyclones made with graphite-lled nylon
eliminate the problem. Use of conductive samplers and lter cassettes (Omega Specialty Instrument
Co., 4 Kidder Road, Chelmsford, MA 01824) is recommended.
2. Precision: The gure 0.068 mg quoted above for the precision is based on a study [3] of weighing
procedures employed in the past by the Mine Safety and Health Administration (MSHA) in which
lters are pre-weighed by the lter manufacturer and post-weighed by MSHA using balances
readable to 0.010 mg. MSHA [14] has recently completed a study using a 0.001 mg balance for the
post-weighing, indicating imprecision equal to 0.006 mg.
Imprecision equal to 0.010 mg was used for estimating the LOD and is based on specic suggestions
[8] regarding lter weighing using a single 0.001 mg balance. This value is consistent with another
study [15] of repeat lter weighings, although the actual attainable precision may depend strongly
on the specic environment to which the lters are exposed between the two weighings.
REFERENCES:
[1] Bartley DL, Chen CC, Song R, Fischbach TJ [1994]. Respirable aerosol sampler performance testing.
Am Ind Hyg Assoc J, 55(11): 1036–1046.
[2] Bowman JD, Bartley DL, Breuer GM, Shulman SA [1985]. The precision of coal mine dust sampling.
Cincinnati, OH: National Institute for Occupational Safety and Health, DHEW (NIOSH) Pub. No.
85-220721.
[3] Parobeck P, Tomb TF, Ku H, Cameron J [1981]. Measurement assurance program for the weighings
of respirable coal mine dust samples. J Qual Tech 13:157.
[4] ACGIH [1996]. 1996 Threshold limit values (TLVs™) for chemical substances and physical agents
and biological exposure indices (BEIs™). Cincinnati, OH: American Conference of Governmental
Industrial Hygienists.
[5] American Conference of Governmental Industrial Hygienists [1991]. Notice of intended change
appendix D—particle size-selective sampling criteria for airborne particulate matter. Appl Occup
Env Hyg 6(9): 817–818.
[6] NIOSH [1977]. NIOSH Manual of sampling data sheets. Cincinnati, OH: National Institute for
Occupational Safety and Health, DHEW (NIOSH) Publication No. 77-159.
[7] Higgins RI, Dewell P [1967]. A gravimetric size selecting personal dust sampler. In: Davies CN, Ed.
Inhaled particles and vapors II. Oxford: Pergammon Press, pp. 575–586.
[8] Bowman JD, Bartley DL, Breuer GM, Doemeny LJ, Murdock DJ [1984]. Accuracy criteria
recommended for the certication of gravimetric coal mine dust personal samplers. NTIS Pub. No.
PB 85-222446 (1984).
[9] Breslin, JA, Page SJ, Jankowski RA [1983]. Precision of personal sampling of respirable dust in coal
mines. U.S. Bureau of Mines Report of Investigations #8740.
[10] Bartley DL, Breuer GM [1982]. Analysis and optimization of the performance of the 10-mm cyclone.
Am Ind Hyg Assoc J 43: 520–528.
[11] Caplan KJ, Doemeny LJ, Sorenson S [1973]. Evaluation of coal mine dust personal sampler
performance, Final Report. NIOSH Contract No. PH CPE-r-70-0036.
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition
PARTICULATES NOT OTHERWISE REGULATED, RESPIRABLE: METHOD 0600, Issue 3, dated 15 January 1998 - Page 5 of 6
[12] Bartley DL, Breuer GM, Baron PA, Bowman JD [1984]. Pump uctuations and their eect on cyclone
performance. Am Ind Hyg Assoc J 45(1): 10–18.
[13] Briant JK, Moss OR [1983]. The inuence of electrostatic charge on the performance of 10-mm
nylon cyclones. Unpublished paper presented at the American Industrial Hygiene Conference,
Philadelphia, PA, May 1983.
[14] Koqut J [1994]. Private Communication from MSHA, May 12, 1994.
[15] Vaughn NP, Chalmers CP, Botham [1990]. Field comparison of personal samplers for inhalable dust.
Ann Occup Hyg 34: 553–573.
METHOD REVISED BY:
David L. Bartley, Ph.D., NIOSH/DPSE/ARDB and Ray Feldman, OSHA.
Figure 1. Bias of three cyclone types relative to the international respirable dust sampling convention.
Aluminum cyclone at 2.5 L/min
Mass median diameter (µm)Mass median diameter (µm)
Higgins-Dewell cyclone at 2.2 L/min
Geometric standard deviation
Mass median diameter (µm)
10-mm Nylon cyclone at 1.7 L/min
APPENDIX: Jarless Method for Calibration of Cyclone Assemblies
This procedure may be used in the eld to calibrate an air sampling pump and a cyclone assembly
without using the one-liter calibration jar.
1. Connect the pump to a pressure gauge or water manometer and a light load (adjustable valve or
5-µm lter) equal to 2” to 5” H
2
O with a TEE” connector and exible tubing. Connect other end of
valve to an electronic bubble meter or standard bubble tube with exible tubing (See Fig. 2.1).
NOTE: A light load can be a 5-µm lter and/or an adjustable valve. A heavy load can be several 0.8-
µm lters and/or adjustable valve.
2. Adjust the pump to 1.7 L/min, as indicated on the bubble meter/tube, under the light load
conditions (2” to 5” H
2
O) as indicated on the pressure gauge or manometer.
3. Increase the load until the pressure gauge or water manometer indicates between 25” and 35” H
2
O.
Check the ow rate of the pump again. The ow rate should remain at 1.7 L/min ± 5%.
4. Replace the pressure gauge or water manometer and the electronic bubble meter or standard
bubble tube with the cyclone having a clean lter installed (Fig. 2.2). If the loading caused by the
cyclone assembly is between 2” and 5” H
2
O, the calibration is complete and the pump and cyclone
are ready for sampling.
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition
PARTICULATES NOT OTHERWISE REGULATED, RESPIRABLE: METHOD 0600, Issue 3, dated 15 January 1998 - Page 6 of 6
Figure 2.1. Block diagram of pump/load/ow meter set-up.
Pressure
gauge
Tee
Air sampling pump
Valve
and/or
filter load
Air flow
Flow meter
Figure 2.2. Block diagram with cyclone as the test load.
Pressure
gauge
Tee
Cyclone filter
Cyclone
Air sampling pump